B-21
B.14 BIS Module Specifications
BIS Module
Type Single-slot
module
Standard IEC60601-2-26
Measurement method
Bispectral index, power spectrum analysis
Measured parameters
EEG
BIS: 0 to 100
Calculated parameters
SQI
EMG
SR
SEF
TP
Impedance range
0 to 999 k
Ω
Sweep speed
6.25, 12.5, 25 or 50 mm/s
Input impedance
>
50 M
Ω
Noise (RTI)
<
0.3 µV (0.25 to 50 Hz)
Input signal range
±1 mV
EEG bandwidth
0.25 to 110 Hz
Patient leakage current
<
10 µA
BIS alarm limits
Range
Step
Unit
BIS High Limit
(low limit + 2) to 100
BIS Low Limit
0 to (high limit – 2)
1 %