13.6.3 Install the BIS Module
13.7 Pneumatic Connectors
This anesthesia machine provides two types of connectors —pipeline connectors (for O
2
,
N
2
O and AIR) and cylinder connectors (for O
2
and N
2
O).
For the pipeline connectors, four types of configuration are available:
O
2
O
2
and N
2
O
O
2
and AIR
O
2
, N
2
O and AIR
For the cylinder connectors, three types of configurations are available:
O
2
O
2
and N
2
O
O
2
and O
2
For details, refer to
2.2 Equipment Appearance
WARNING
z
Use medical grade gas supplies only. Other types of gas supplies may contain
water, oil, or other contaminants.
z
When the central piping system fails, one or more equipment connected may stop
work. Make sure that cylinders are available.
z
When gas supplies are cut off, there is still pressure inside the pipeline. Remember
to release the gas inside the pipeline before removing the tube.
z
If the [Drive Gas Pressure Low] alarm occurs when the gas supply pressure is
greater than 200 kPa, contact your service personnel or us.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...