6-12
6.9 Alarm Tests
The anesthesia machine performs a self test after started. The alarm lamp flashes yellow and
red once in turn and then a beep is given. Then the display shows the start-up screen and
enters the standby screen after 30 seconds. This means that audio and visual alarm indicators
begin to work normally.
6.9.1 Prepare for Alarm Tests
1. Connect a test lung or manual bag to the Y piece patient connection.
2. Set the bag/mechanical ventilation switch to the mechanical (
) position.
3. Set the system switch to the
position.
4. Set the system to Standby.
5. Set the ventilator controls as follows:
Ventilation mode: select the [
Vent Mode
] shortcut key and then [
VCV
].
[
TV
]: 500 ml.
[
Rate
]: 12
BPM.
[
I:E
]: 1:2.
[
Plimit
]: 30 cmH
2
O.
[
PEEP
]: OFF.
6. Push the O
2
flush button to fill the bellows, folding bag rising to the top.
7. Turn the O
2
flow control to set the O
2
flow to 0.5 to 1 L/min.
8 Press
the key and select [
Ok
] from the pop-up menu to exit the standby status.
9. Make sure that:
The ventilator displays the correct data.
The folding bag inside the bellows inflates and deflates normally during
mechanical ventilation.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...