14-2
NOTE
z
Do not soak synthetic rubber parts for more than 15 minutes. Swelling or faster
aging can occur.
z
Only autoclave parts marked 134
º
C.
z
Cleaning solutions must have a pH of 7.0 to 10.5.
14.1 Clean and Disinfect the Anesthesia Machine
Housing
1. Clean the surface of the anesthesia machine housing with a damp cloth soaked in mild
detergent (such as 70% ethanol).
2. After cleaning the housing, remove the remaining detergent by wiping with a dry lint
free cloth.
WARNING
z
Seeping liquid into the control assembly can damage the equipment or cause
personal injury. When cleaning the housing, make sure that no liquid flows into the
control assemblies and always disconnect the equipment from the AC mains.
Reconnect the AC mains after the cleaned parts are fully dry.
NOTE
z
Use only soft dry and lint free cloth to clean the display. Do not use any liquid for
display cleaning.
14.2 Disassemble the Breathing System Cleanable
Parts
You need to disassemble the breathing system cleanable parts first before cleaning the
system.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...