6.8.3 Breathing System Leak Test in Manual Ventilation Status
1. Make sure that the system is Standby. If not, press the
key and select [
Ok
] from the
pop-up menu to enter standby status.
2. Set the bag/mechanical ventilation switch to the bag position.
3. Connect the manual bag to the manual bag port.
4. Turn the APL valve control to fully close the APL valve (75 cmH
2
O).
5. Turn the O
2
flow control to set the O
2
flow to 0.15-0.2 L/min.
6. Connect the Y piece on the breathing tube to the leak test plug on the manual bag port.
Occlude the gas outlet of the Y piece.
7. Push the O
2
flush button to let the pressure increase to approximately 30 cmH
2
O on the
airway pressure gauge.
8. Release the flush button. A pressure decrease on the airway pressure gauge indicates a
leak. Look for and please contact your service personnel.
6.8.4 APL Valve Test
1. Make sure that the system is Standby. If not, press the
key and select [
Ok
] from the
pop-up menu to enter Standby.
2. Set the bag/mechanical ventilation switch to the bag position.
3. Connect the manual bag to the manual bag port.
4. Connect the Y piece on the breathing tube to the leak test plug on the manual bag port.
5. Turn the APL valve control to let the pressure of APL valve stay at 30 cmH
2
O.
6. Push the O2 flush button to inflate the manual bag.
7. Make sure that the reading on the airway pressure gauge is with the range of 20 to 40
cmH
2
O.
8. Turn the APL valve control to the MIN position.
9. Set the O
2
flow to 3 L/min. Turn any other gases off.
10. Make sure that the reading on the airway pressure gauge is less than 5 cmH
2
O.
11. Push the O
2
flush button. Make sure that the reading on the airway pressure gauge does
not exceed 10 cmH2O.
12. Turn the O
2
flow control to set the O
2
flow to minimum. Make sure that the reading on
the airway pressure gauge does not decrease below 0 cmH
2
O.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...