4-23
4.7 Stop Mechanical Ventilation
To stop mechanical ventilation, do as follows:
1. Make sure that the breathing system is set up and the APL valve is set properly before
stopping mechanical ventilation.
The APL valve adjusts the breathing system pressure limit during manual ventilation. Its
scale shows approximate pressure.
2. Set the bag/mechanical ventilation switch to the
position. This selects manual
ventilation and stops mechanical ventilation (ventilator).