6-16
WARNING
z
Before connecting a patient, flush the anesthesia machine with 5 L/min of O
2
for at
least one minute. This removes unwanted mixtures and by-products from the
system.
6.11 Inspect the AGSS
Assemble the AGSS as described in
13.10.2Assemble the AGSS
and then turn on the waste
gas disposal system. Check if the float can rise and exceed the “MIN” mark. If any blockage,
tackiness, or damage occurs to the float, disassemble and assemble the float again or replace
the float.
NOTE
z
Do not block the AGSS pressure compensation openings during the inspection.
If the float cannot rise, the possible reasons are:
1. The float is tacky. Turn over the AGSS and check if the float moves up and down freely.
2. The float is rising slowly. The filter may be blocked. Check if the filter is blocked as
described in
14.2.13.1Filter.
3. The waste gas disposal system is not working or the pump rate is less than 50 L/min at
which the AGSS works normally. Check the waste gas disposal system as described in
13.10.3 Waste Gas Disposal System.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...