4-21
4.4.6.6 Parameter Range and Default Value in PSV Mode
Parameter
Range
Step
Default
Ventilation mode
Pinsp
5 to 60 cmH2O
1 cmH2O
15 cmH2O
Rate
4 to 60 BPM
1 BPM
10 BPM
I:E
4 to 100 BPM
1 BPM
12 BPM
PCV (backup
ventilation mode)
Finsp
20 to 85 L/min
1 L/min
60 L/min
Plimit
10 to 100 cmH2O
1 cmH2O
30 cmH2O
Psupp
5 to 60 cmH2O
1 cmH2O
15 cmH2O
PEEP
OFF, 4 to 30
cmH2O
1 cmH2O
OFF
Backup Mode Active 5 to 30 s
5 s
30 s
Pressure
-20 to -1 cmH2O
1 cmH2O
-2 cmH2O
Trigger
Level
Flow
0.5 to 15 L/min
0.5 L/min
3 L/min
PSV Insp
Termination Level
5 to 60 %
5 %
25 %
PSV
4.5 Start Mechanical Ventilation
After settings of the related parameters are already made, you can enter mechanical
ventilation mode by pushing the Standby key
on the panel and then selecting [
Ok
] from
the pop-up menu to exit the standby status. The system will then work in the selected
mechanical ventilation mode.
NOTE
z
Before starting a new mechanical ventilation mode, make sure that all related
parameters are set appropriately.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...