B
-8
Rate
4 to 100 BPM
1 BPM
VCV, PCV, PSV
I
:
E
4:1 to 1:8
0.5
VCV, PCV, PSV
TIP:TI
OFF, 5 to 60%
5%
VCV
Finsp
20 to 85 L/min
1 L/min
PSV
、
SIMV-VC
、
SIMV-PC
Trigger Window
5 to 90%
5%
SIMV-PC, SIMV-VC
SIMV Rate
4 to 60 BPM
1 BPM
SIMV-VC,SIMV-PC
Tinsp
0.4 to 5 s
0.1 s
SIMV-VC, SIMV-PC
Inspiratory Trigger
Level
Pressure:
PEEP-20 cmH
2
O to
PEEP-1 cmH
2
O
Flow:
0.5 to 15 L/min
Pressure:
-1 cmH
2
O
Flow:
0.5 L/min
PSV
、
SIMV-VC
、
SIMV-PC
PSV Insp Termination
Level
5 to 60%
5%
PSV, SIMV-VC,
SIMV-PC
Backup Mode Active
5 to 30 s
5 s
PSV
PEEP setting range
Type
Integrated electronic PEEP
Range
OFF, 4 to 30 cmH
2
O; increment:1 cmH
2
O
Ventilator performance
Drive pressure
280 to 600 kPa
Peak flow
100 L/min
Range of flow valve
1 to 100 L/min
Ventilator monitored parameters
MV
0 to 100 L/min
TV
0 to 2500 mL
O2 concentration
18 to 100%
Paw
-20 to 120 cmH
2
O
Pmean
-20 to 120 cmH
2
O
Pplat
-20 to 120 cmH
2
O
I
:
E
4:1 to 1:10
PEEP monitored parameter
Range
0 to 70 cmH
2
O
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...