13-22
2. Push and turn the locking lever A clockwise to lock the vaporizer in position.
3. Make sure that the top of the vaporizer is horizontal. If not, remove the vaporizer and
reinstall it.
4. In case of reinstalling the vaporizer, try to lift each vaporizer straight up off the manifold
rather than pulling forward. Do not rotate the vaporizer on the manifold.
5. If a vaporizer lifts off the manifold, install it again and complete steps 1 through 3. If the
vaporizer lifts off a second time, do not use the system.