14-16
All parts of the breathing system can be cleaned and disinfected. The cleaning and
disinfection methods are different for different parts.
You need to select the appropriate method to clean and disinfect the parts based on the actual
situations to avoid cross-contamination.
This table is our recommended cleaning and disinfection methods for all parts of the
breathing system.
Intermediate level disinfection
High level
disinfection
Parts
A* B* C*
Breathing tubes and Y piece
★
★
Breathing mask
★
★
Flow sensor
★
Bellows assembly
★
★
Inspiratory and expiratory check
valves assemblies
★
★
O
2
sensor
★
Canister assembly
★
★
Canister connection block
assembly
★
★
Water collection cup
★
★
Bag arm
★
★
BYPASS assembly
★
★
Breathing system
★
★
Manual bag
★
★
AGSS assembly
★
★
indicates that this disinfection method is applicable.
A*. Clean with a damp cloth soaked in mild detergent and then wipe off the remaining
detergent with a dry lint free cloth.
B*. Flush with water first; then soaked in water and cleaning solution (water temperature
40ºC recommended) for approximately three minutes and wipe with 70% ethanol.
C*. Steam autoclave at maximum 134ºC.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...