8-7
8.3.5 Scavenge the Sample Gas
To scavenge the sample gas to the waste gas disposal system, depress the metal chip and then
plug the exhaust tube to the ports marked
(sample gas return to the AGSS) on the
anesthesia machine as shown in the above picture.
WARNING
z
When using CO2 module to perform CO2
measurements on the patient who is
receiving or has recently received anesthetic agents, connect the gas outlet to the
waste gas disposal system to prevent the medical staff from breathing in the
anesthetic agent.
8.3.6 Zero the Sensor
Zeroing the sensor aims to eliminate the effect of baseline drift on the readings during the
measurement so as to ensure measurement accuracy.
For CO2 module, a zero calibration is carried out automatically when necessary. You can also
start a manual zero calibration when deemed necessary. To manually start a zero calibration,
enter the [
Gas Module Setup >>
] menu and then select [
Zero
]. You do not need to
disconnect the sensor from the breathing system when performing the zeroing.
8.3.7 Calibrate the Sensor
For CO2 module, a calibration should be performed once a year or when the measured value
has a great deviation.
Metal chip
Exhaust tube
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...