5-13
5.3.7 Volume Monitoring
5.3.7.1 Display Volume Waveform
5.3.7.2 Set Volume Waveform
1. Select the waveform area to access the waveform setup
menu.
2. Select
[
Waveform
] and select [
Volume
].
3. Select
[
Sweep
] and toggle between [
6.25 mm/s
] and [
12.5 mm/s
]. The greater the value
is, the faster the waveform sweeps.
4. Select
[
Scale
] and toggle between [
500
], [
1000
], and [
1500
].The volume ranges
corresponding to the waveform scales are:
[
500
]: 0 to 500 ml.
[
1000
]: 0 to 1000 ml.
[
1500
]: 0 to 1500 ml.
5. Select
to exit the current menu.
5.3.8 Breath Rate Monitoring
5.3.8.1 Display Breath Rate
Refer to
5.3.5.1Display Tidal Volume and Breath Rate Parameters
.
5.3.8.2 Set Breath Rate Alarm Limits
1. Select
the
[
Alarm Setup
] shortcut key and select [
Ventilator >>
].
2. Set Rate high and low alarm limits in the [
Ventilator Alarm Limits
] menu.
3 Select to exit the current menu.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...