13-33
WARNING
z
The anesthesia machine stops gas delivery when the supply gas pressure is lower
than 200 kPa.
13.7.1 Connect the Pipeline Gas Supplies
The anesthesia machine provides three (O2, N
2
O and AIR) pipeline supply connectors which
are connected to three tubes of different colors and cannot be exchanged. Connect the
pipeline gas supplies as follows:
1. Check that the seal at the tube connector is in good condition before connecting the gas
supply tube. If damaged, do not use the tube. Replace the seal to avoid leakage.
2. Align the tube connector with the matching gas supply connector at the back of the
anesthesia machine and then insert it.
3. Make sure that the tube is properly connected and tighten the tube nut.
Seal