13-34
13.7.2 Install the Gas Cylinder
For details, refer to
13.5Install/Replace the Gas Cylinder
13.8 CIS Connector
The anesthesia machine can be connected to an anesthesia information system (CIS), which
is to be installed, serviced and updated by Mindray authorized or approved personnel. For
details, refer to the Instructions for Use accompanying the CIS.
13.9 Scavenging
The scavenging assembly is located on the left side of the work table. There are two outlets
labeled AGSS and PEEP as shown below.
1. The PEEP outlet gives off the exhaust gas indoors directly.
2. The outside diameter of the AGSS connector is 30 mm. with 1:20 taper ratio. Please
connect to the AGSS or waste gas disposal system.
WARNING
z
The PEEP outlet gives off a small amount of O2 continuously. Do not occlude this
outlet. Otherwise, the anesthetic ventilator cannot work normally.
z
Before performing an operation on the patient, equip the anesthesia machine with
anesthesia gas scavenging system which complies with ISO 8835-3 to purify the air
in the operating room.
z
If your anesthesia machine is not configured with active AGSS, do not connect the
waste gas exhaust port of the anesthesia machine to the active hospital’s waste gas
disposal system.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...