B-2
B.2 Environmental Specifications
Main unit
Item
Temperature (ºC)
Related humidity
(non-condensing)
Barometric pressure
(kPa)
Operating
10 to 40
15 to 95%
70 to 106
Transport storage
–20 to +55
10 to 95%
50 to 106
AG module
Item
Temperature (ºC)
Related humidity
(non-condensing)
Barometric pressure
(kPa)
Operating
10 to 40
15 to 95%
70 to 106
Transport storage
–20 to +55
10 to 95%
70 to 106
B.3 Power Requirements
External AC power supply
Input voltage
100 to 240 V
100 to 120 V
220 to 240 V
Input current
8.5 to 3.5 A
8.5 A
3.5 A
Input frequency
50/60 Hz
Leakage current
< 500
μ
A
Fuse T10
AL/250V
Power cord
5 m
Auxiliary output supply (with isolation transformer)
Output voltage
220 to 240 V
100 to 120 V
Output frequency
50/60 Hz
50/60 Hz
Output current
(
outlet 1
)
1.6A
3.8A
Output current
(
outlet 2
)
0.5A
1.0A
Output current
(
outlet 3
)
0.5A
1.0A
Fuse
(
outlet 1
)
T
3.0AL/250V
T6.3AL/250V
Fuse
(
outlet 2
)
T1.6AL/250V
T1.6AL/250V
Fuse
(
outlet 3
)
T1.6AL/250V
T1.6AL/250V
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...