6-7
NOTE
z
When O
2
supply is disconnected, alarms for [O2 Supply Failure] and [Drive Gas
Pressure Low] occur as O
2
pressure decreases.
9. Set the system switch to the
position.
6.6.2 With O2 Sensor
Do as described in
6.9.2 Test the O2 Concentration Monitoring and Alarms
before testing.
To do the flow control system tests:
1. Connect the pipeline supplies or slowly open the cylinder valves.
2. Turn all flow controls fully clockwise (minimum flow).
3. Set the system switch to the
position.
4. Do not use the system if low battery or other ventilator failure alarms occur.
5. Adjust all gas flows to minimum.
Steps 6 and 7 are only for systems with N
2
O.
WARNING
z
During steps 6 and 7, the O
2
sensor used must be correctly calibrated and the Link
system should be kept engaged.
z
Adjust only the test control (N
2
O in step 6 and O
2
in step 7).
z
Test the flows in sequence (N
2
O then O
2
).
6. Test the O
2
-N
2
O Link system with flow increasing:
Turn the N
2
O and O
2
flow controls fully clockwise (minimum flow).
Slowly turn the N
2
O flow control counterclockwise.
Make sure that the O
2
flow increases. The measured O
2
concentration must be
≥
21% through the full range.
7. Test the O
2
-N
2
O Link system with flow decreasing:
Turn the N
2
O flow control and set the N
2
O flow to 9.0 L/min.
Turn the O
2
flow control and set the O
2
flow to 3 L/min or higher.
Slowly turn the O
2
flow control clockwise.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...