4-11
4.4.4.6 Parameter Range and Default Value in PCV Mode
Parameter
Range
Step
Default
Pinsp
5 to 60 cmH2O
1 cmH2O
15 cmH2O
Rate
4 to 100 BPM
1 BPM
12 BPM
I:E
4:1 to 1:8
0.5
1:2
Plimit
10 to 100 cmH2O
1 cmH2O
30 cmH2O
PEEP
OFF, 4 to 30 cmH2O
1 cmH2O
OFF
4.4.5 Synchronized Intermittent Mandatory Ventilation (SIMV)
This anesthesia machine supports two modes of SIMV: SIMV-volume control (SIMV-VC)
and SIMV–pressure control (SIMV-PC).
4.4.5.1 Description
SIMV-VC
SIMV-VC means to deliver volume controlled breathing to the patient by phase at the preset
intermission. In the SIMV-VC mode, the ventilator waits for patient’s next inspiration based
on the specified time interval. The sensitivity depends on [
Trigger Level
] (optional flow and
pressure). If [
Trigger Level
] is reached within the trigger waiting time (called synchronous
[
Trigger Window
]), the ventilator delivers volume controlled breathing synchronously with
the preset tidal volume and inspiratory time. If the patient does not inspire within the
[
Trigger Window
], the ventilator delivers volume controlled breathing to the patient at the
end of [
Trigger Window
].Spontaneous breathing outside of [
Trigger Window
] can acquire
pressure support.
SIMV-PC
SIMV-PC means to deliver pressure controlled breathing to the patient by phase at the preset
intermission. In the SIMV-PC mode, the ventilator waits for patient’s next inspiration based
on the specified time interval. The sensitivity depends on [
Trigger Level
] (optional flow and
pressure). If [
Trigger Level
] is reached within the trigger waiting time (called synchronous
[
Trigger Window
]), the ventilator delivers pressure controlled breathing synchronously with
the preset tidal volume and inspiratory time. If the patient does not inspire within the
[
Trigger Window
], the ventilator delivers pressure controlled breathing to the patient at the
end of [
Trigger Window
].Spontaneous breathing outside of [
Trigger Window
] can acquire
pressure support.
If [
Trigger Level
] is reached outside of [
Trigger Window
], the ventilator delivers
pressure-supported ventilation based on the preset [
Psupp
].
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...