10-7
1. Measure
electrode
impedance
2. Time of the most recent impedance check
3. Start/stop cyclic impedance checks
4. Show sensor information
The measured electrode-to-skin impedance and electrode status are displayed above each
electrode:
Status
Description
Action
[
Lead off
]
Electrode falls off and has no skin
contact.
Reconnect electrode, or check the
sensor-to-skin contact. If necessary,
clean and dry skin.
[
Noise
]
The EEG signal is too noisy.
Impedance cannot be measured.
[
High
]
The impedance is above the limit.
Check the sensor-to-skin contact. If
necessary, clean and dry skin.
[
Pass
]
The impedance is within valid range.
No action necessary.
Although BIS may still be measured when the electrode status is [
Noise
] or [
High
], all
electrodes should be in [
Pass
] status for the best performance.
10.9 Set BIS Smoothing Rate
Select [
Smoothing Rate
] from the [
BIS
Module
Setup
>>] menu and toggle between [
10
s
],
[
15 s
] and [
30 s
].
The smoothing rate defines how the anesthesia machine averages the BIS value. A smaller
smoothing rate indicates increased responsiveness to changes in the patient’s state. A bigger
smoothing rate indicates a smoother BIS trend with decreased sensitivity to artifacts.
10.10 Restore Defaults
Select [
Defaults
] from the [
BIS Module Setup
] menu. Then all the options in this menu
except [
Cont
.
Imped
.
Check
] and [
Cyc
.
Imped
.
Check
] are restored to the factory default
configurations.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...