1.6.9
Sequence for Establishing SpO
2
with Masimo® Pulse Oximetry
1.
Select the appropriate sensor for the patient from the table below. All sensors below are
non-sterile and can be used during patient movement.
2.
Attach the PC12 Patient Cable (P/N 0012-00-1099-02) to the sensor and plug the other
end of the patient cable into the SpO
2
connector (28) located on the left side panel of
the monitor.
*
This feature applicable only if available / installed in your unit.
NOTE:
The PC12 Patient Cable is not used with the LNOPDCSC
Sensors.
NOTE:
Do not place the sensor on an extremity with an invasive
catheter or blood pressure cuff in place.
NOTE:
Ensure proper routing of patient cable to avoid
entanglement and/or strangulation.
WARNING: When equipped with Nellcor® SpO
2
, use only Nellcor®
oxygen transducers including Nellcor® OXISENSOR™
patient dedicated adhesive sensors. Use of other oxygen
transducers may cause improper oximeter performance.
WARNING: Tissue damage or inaccurate measurements may be caused
by incorrect sensor application or use, such as wrapping it
too tightly, applying supplemental tape, failing to inspect
the sensor site periodically, or failing to position it
appropriately. Carefully read the sensor directions for use,
the Passport operating instructions, and all precautionary
information before use.
WARNING: Excessive ambient light may cause inaccurate
measurements. Cover the sensor site with opaque material.
*
TABLE 1-1 Masimo® Sensor Family
SELECTION
PART NUMBER
PATIENT SIZE
DISPOSABLE
/ REUSABLE
LNOP® - Adt Adult Disposable Finger
Sensor
0600-00-0043-01
> 30 kg.
Disposable
LNOP® -Pdt Pediatric/Slender Digit
Disposable Sensor
0600-00-0044-01
10 to 50 kg.
Disposable
LNOP® - Neo Neonatal Disposable
Sensor
0600-00-0045-01
> 2000 gm.
Disposable
LNOP® - Neo Pt Neonatal Pre-term
Disposable Sensor
0600-00-0046-01
< 2000 gm.
Disposable
LNOP® - DC1 Adult Reusable Finger
Sensor
0600-00-0047
> 30 kg.
Reusable
LNOP® DCSC Adult Reusable Spot
Check Sensor
0600-00-0077
> 30 kg.
Reusable
LNOP® YI Multisite Reusable Sensor
0600-00-0078
> 1 kg.
Reusable
LNOP® EAR Reusable Ear Sensor
0600-00-0079
> 30 kg.
Reusable
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...