Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 107
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
NOTE:
The sample display shown may not be an exact
representation of your unit.
B. Trend Lists
FIGURE 1-63
Trend List Display
Pressing the DISPLAY key (19) displays the trend lists, as shown above. The trend lists take up
the entire Multi-function area. To view additional parameters press the DISPLAY key (19)
again. Pressing this key a third time returns the screen to normal. Pressing and holding the
DISPLAY key (19) for 3 seconds will clear all trended data.
Each entry to the trend list is added to the bottom of the list until the page is full. Nine lines of
data can be displayed on each page. The trend data memory is large enough to contain
more than one page of data. When the most current page is full and new data is available,
the top line of data in the window is removed (but kept in memory). The other lines of data
are scrolled up and the new data is added as the last line of data in the window. To scroll
through the data that is in memory press the
q
and
p
trend keys (20).
The trend lists display consists of a heading that includes the page #, the time, the
parameters with their units, and the listed data in columns under the appropriate heading.
If no data is available for a particular column, dashes (—-) are displayed. If an NIBP
measurement was attempted and a valid reading was unable to be obtained (xxx) will be
displayed in the NIBP column.
If respiration is turned off, “OFF” is displayed in the respiration column.
NOTE:
The IBP1, CO
2
and Gas Module columns display only in
models that are equipped with these options. If CO
2
is being
monitored, the CO
2
trend will replace the Temp trend. All
Gas Module trends will be on a separate page.
The Trend Trigger and Interval are set in the User Configuration Trend menu. See “User
Configuration Mode” on page 1-153.
To clear the trend data, press and hold the DISPLAY key (19) for 3 seconds.
NOTE:
Trend data will be stored in memory for 1 hour after the
unit has been powered down. After 1 hour the trend data
will be cleared.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...