Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 181
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
1.6.23.8
SpO
2
Menu / Nellcor SpO
2
/ Masimo SpO
2
Menu
Pleth Size:
Sets the scale of the plethysmograph waveform.
Sensor-Off Audio:
Set to Off - no beep tone from the Passport. Set to ON - a one time
series of 5 triple beeps upon recognition of a SpO
2
sensor off condition. Set to CONT - a
series of 5 triple beeps every 30 seconds.
Mode (Nellcor SpO
2
only):
Sets the averaging time of the plethysmograph waveform on
units with Nellcor. Scaling is automatic, refer to “Special Features” on page 1-131.
Sensitivity Mode (Masimo SpO
2
only):
Sets the sensitivity mode depending upon
signal quality.
Post Average Time (Masimo SpO
2
only):
Changes averaging time for Saturation,
Pulse Rate, and Signal Strength measurements for SpO
2
.
FIGURE 1-100
SpO
2
Menu
1.6.23.9
CO
2
Menu
CO
2
Scale:
Select the size of the displayed CO
2
waveform.
Start Zero Calibration:
Select Yes to initiate the calibration of the sensor.
Start adapter cal:
Select Yes to initiate the calibration of the airway adapter with the
sensor.
CHANGE SPO2
Pleth Size:
Choices: 1, 2, 3, 4
= Adjust valve
EXIT = quit
SELECT = Enter/move
3
CHANGE SPO2
Mode:
Choices: 1, 2, 3
= Adjust valve
EXIT = quit
SELECT = Enter/move
1
Sensor-Off Audio:
ON
CHANGE SPO2
Pleth Size:
Choices: ON, OFF, CONT
= Adjust valve
EXIT = quit
SELECT = Enter/move
3
Sensor-Off Audio:
ON
Sensor-Off Audio:
ON
CHANGE SPO2
Mode:
Choices: ON, OFF, CONT.
= Adjust valve
EXIT = quit
SELECT = Enter/move
1
Sensor-Off Audio:
ON
Datascope SpO2
Nellcor SpO2
CHANGE SPO2
Choices: ON, OFF, CONT
= Adjust valve
EXIT = quit
SELECT = Enter/move
ON
:
8
CHANGE SPO2
Sensitivity Mode: Normal
Choices: 4,6,8,10,12,14,16
= Adjust valve
EXIT = quit
SELECT = Enter/move
ON
Post Average Time (sec)
8
Masimo SpO2
Sensor-Off Audio
Post Average Time (sec)
Sensor-Off Audio:
Sensitivity Mode: Normal
CHANGE SPO2
Choices: Normal, High
= Adjust valve
EXIT = quit
SELECT = Enter/move
ON
:
8
Sensor-Off Audio
Post Average Time (sec)
Sensitivity Mode:
Normal
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...