Cable Assembly Details
Repair Information
4 - 26
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
HP ECG/ECG Interface - 0012-00-0753-01 (A)
HP ECG/ECG Interface (XG-CE)- 0012-00-1027 (A)*
*
This version cable is exactly the same as the other except it also has a ferrite on it for CE approval.
A
See CONNECTOR
.870 Dia
View A-A
B
E
C
D
F
J
H
A
H
5
9
4
N.C.
AGND
ECG
AGND
P1
P2
1 - 3
6 - 8
10 - 16
N.C.
N.C.
N.C.
All other pins
10 k
10
B
D
SEE NOTE 1
NOTE:
1. CABLE SHIELD SHALL NOT CONTACT CONNECTOR SHIELD OR FERRULE.
8',+/- .5'
Terminal 1
P1
P2
A
A