Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 118
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
1.6.6.2
Neonatal Electrode Placement
FIGURE 1-71
Neonatal Electrode Placement
Lead placement on the neonate with a three lead set is directed towards obtaining the best
respiration waveform. The thoracic impedance is measured between the RA and LA
electrodes therefore, the electrodes must be placed across the chest from each other to
optimize measurement of chest movement.
1.6.7
Invasive Pressure Acquisition
1.
Plug the pressure transducer into the PRESSURE TRANSDUCER connector (31) or (32) on
the side panel.
• Place lead select on lead II.
• Select lead I for monitoring.
• For MCL, select lead II for monitoring.
FIGURE 1-69
3 Lead Electrode
Placement (standard
configuration)
FIGURE 1-70
3-Lead Electrode
Placement (for MCL)
• Place black electrode on left shoulder
under clavicle.
• Place white electrode on left shoulder
under clavicle.
RA
LA
LL
Red
W hit e
Blac k
L L
L A
R A
W hite
Red
Bl ac k
RA
LA
G
Red
Black
White