Warnings, Precautions and Notes
Introduction
Cautions
CAUTION:
Connection of non-isolated devices to the optional RS232
Connector on this unit may cause chassis leakage to exceed
the specification standards.
CAUTION:
Only use the Abbreviated Operating Check List if you are
already familiar with this product. If not, please continue
with the remainder of this chapter, Detailed Operating
Instructions.
CAUTION:
Always attach all desired peripheral equipment prior to
powering on the Passport. If required to attach more or
change what is connected, turn the Passport off, attach
what is needed, and then turn the Passport on again.
CAUTION:
Observe extreme caution on all patients (neonates,
pediatrics, and adults) when NIBP is set to the Continuous
Mode. A 5 minute limit is placed on continuous
measurements. After 5 minutes the unit will take a
measurement once every 5 minutes. Reports have been
made of nerve injury occurring during use of automatically
cycled blood pressure cuffs. See the Appendix, “Precautions
when Using Automatically Cycled Blood Pressure
Cuffs”.Observe extreme caution on all patients (neonates,
pediatrics, and adults) when NIBP is set to the Continuous
Mode. A 5 minute limit is placed on continuous
measurements. After 5 minutes the unit will take a
measurement once every 5 minutes.
Reports have been made of nerve injury occurring during
use of automatically cycled blood pressure cuffs. See the
Appendix, “Precautions when Using Automatically Cycled
Blood Pressure Cuffs”.
CAUTION:
It is the users responsibility to confirm correct operation of
the Passport with external devices. It is essential that a
biomedical engineer verify correct operation of the Passport
with an external device. Testing should include verification
that the maximum delay for cardioversion synchronization
does not exceed 60 ms (monitor and defibrillator). Cable
should be labeled with the devices they have been tested
for use with.
CAUTION:
Do not use this special “Y” shaped power cable for any
devices other than the Passport and Gas Module II.
CAUTION:
Vacuum (negative pressure) should not exceed 1 mmHg at
the Passport Pump Exhaust fitting (39). Excessive scavenge
vacuum may result in an “OCCLUSION” message or
damage to the Passport’s internal pump. The scavenge
system if used must be on during calibration.
CAUTION:
It is the users responsibility to confirm correct operation of
the Passport XG with external devices. Before using this
cable it is essential that a biomedical engineer verify correct
operation of the Passport XG with an external device using
this cable. Testing should include verification that the
maximum delay for cardioversion synchronization does not
exceed 60 ms (monitor and defibrillator). Cable should be
labeled with the devices they have been tested for use with.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...