Operation (5-Lead, 5L, LT)
Operation (5-Lead, 5L, LT and XG)
1 - 58
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
1.3.16.1
NIBP Measurement Messages
• Always have an alternate method of BP verification available.
• On vasoconstricted patients, failure to evacuate air from the cuff can distort BP
measurement.
• Do not place cuff on extremity that has an IV.
MESSAGE
REASON
SOLUTION
NIBP: IDLE
Displayed while system is idle.
Press START to take a single
measurement. Press
INTERVAL to select an
interval and start timed
measurements.
NIBP: DEFLATE
Displayed while the timer
mode is stopped, usually by
pressing the DEFLATE key.
Press START to take an
immediate measurement
and resume timed
measurements.
NIBP: INTERVAL
Displayed in the interval
between two timed
measurements.
Press DEFLATE to suspend
timed measurements. Press
INTERVAL to change timer
to OFF to stop timer.
NIBP: INITIALIZING
The NIBP system is being
initialized.
Wait until initialization is
complete, indicated by
NIBP: IDLE message.
Initialization may take up to
3 minutes from power up.
NIBP: FAILURE
The NIBP system has detected
a potential error.
Power cycle unit. If message
reappears, call Service.
NIBP:MEASURING CUFF
Displayed during a
measurement to show the cuff
pressure.
Press DEFLATE to suspend a
measurement and deflate
the cuff.
NIBP:RETRY MOTION
Measurement has been
attempted but no reading was
possible due to detected
motion and the retry limit has
not been reached.
Retry will be attempted.
Have patient remain still.
NIBP:RETRY PUMP HIGHER
A measurement has been
attempted but no reading was
possible due to pulse
detection at highest cuff
pressure and the retry limit has
not been reached.
Retry will be attempted.
Check that appropriate
patient size is set. Preset
initial inflation pressure.
NIBP:RETRY
A measurement has been
attempted but no reading was
possible.
Retry will be attempted.
Check for leaks and quality
of peripheral pulses.
Decrease patient movement.
Switch cuff to another limb.
NIBP:UNABLE TO MEASURE*
An unsuccessful measurement
cycle has been completed.
Switch cuff to another limb.
Decrease patient movement.
Press START to retry. Be
prepared to auscultate BP
manually.
NIBP:CUFF OVERPRESSURE
The internal hardware cuff
pressure check valve has been
tripped.
Power cycle unit. If message
reappears, call Service.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...