Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 164
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
1.6.19.5
Sidestream CO
2
Messages (on Passports with Sidestream CO
2
only)
1.6.19.6
Gas Module Messages (on Passports with Gas Module installed only)
MESSAGE
REASON
SOLUTION
CO
2
: PUMP ON
The CO
2
sidestream pump is on.
Message will display at
all times when the CO
2
sidestream pump is on.
CO
2
: PUMP CAL IN
PROGRESS
The CO
2
sidestream pump is turned on and a
pump calibration has been requested.
Message will clear
when cal is complete.
CO
2
: BREATHS
DETECTED...RETRY
Breaths are detected while the CO
2
sidestream pump calibration is in progress or
an adaptor calibration is in progress.
Move the sensor/
adapter away from any
source of CO
2
.
CO
2
: PUMP CAL
FAILURE
The CO
2
sidestream pump calibration was
not completed successfully. The pump
calibration could have been requested when
the sampling line was not attached correctly.
Check sampling line for
correct assembly. Retry
calibration. If message
persists refer to
technical service
personnel.
CO
2
: PUMP FAILURE
The CO
2
sidestream pump failed.
Refer to technical
service personnel.
CO
2
: OCCLUSION-
CHECK SYSTEM
The CO
2
hardware is indicating the sampling
pump line is blocked while the CO
2
sidestream pump is on.
Check sampling line
and filter for blockage,
clear sampling line if
possible. Replace
sampling line and/or
filter if necessary. If
sampling cell is
replaced, perform a
pump calibration.
CO
2
: NO SSCO
2
ON NEONATE
The patient size is set to Neonate and the
CO
2
sidestream pump is on, or when the
patient size is set to Neonate and the user
presses the CO
2
pump key.
Do not have the patient
size set to Neonate and
press the CO
2
pump
key.
CO
2
: PUMP CAL
COMPLETE
The pump calibration has been successfully
completed.
Message will clear
automatically.
MESSAGE
REASON
SOLUTION
GM: WARMING UP
Appears when the system has been turned on,
and the sensors have not reached their stable
operating temperature.
Wait for the message to
go away. It takes up to
two minutes for the
device to warm up.
GM: AGENT
WARMING UP
This message appears after the GM:
WARMING UP message disappears. It
indiactes that the Agent ID Bench is warming
up and readings will not be available.
Wait for the message to
go away. It takes ups to
five minutes from power-
up for the Agent ID
Bench to warm up.
GM: EXHAUST
BLOCKED
Appears when the system detects a blockage
at the exhaust gas outlet, as indicated by an
increase in internal pressure.
Remove waste gas
scavenging assembly,
check if message
disappears. Check
exhaust line for
blockage and clear if
possible. If message
persists call for service.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...