Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 93
Operation (5-Lead, 5L, LT and XG)
Controls, Indicators and Connectors (XG)
1.5.6
Gas Module (Optional)- Front Panel
FIGURE 1-57
Gas Module Front Panel
NOTE:
This Manual refers to the Gas Module II and the Gas Module
SE as Gas Module. All illustrations represent the Gas Module
SE. Any parts or descriptions unique to either product will be
identified as such.
41. Input Port
This port is used to connect the sampling tubing to the Gas Module.
42. Water Trap Assembly (includes Water Trap Reservoir)
The Water Trap Assembly (P/N 0202-00-0129 and 0202-00-0134) is used to capture
moisture drawn in with the patient sample. The Water Trap Reservoir must be emptied and
rinsed (with water only) whenever more than half full or whenever changing patients. To
remove push latch to left. Refer to section 4.11 of the Passport Operating Instructions for
more details.
43. Dust Filter
The Dust Filter (P/N 0378-00-0040) protects the Gas Module from airborne dust. It should
be removed and cleaned on a regular basis. Refer to section 4.11 of the Passport Operating
Instructions for more details.
44. Dust Filter Cover
The Dust Filter Cover is removed to access the filter.
SET UP
CENTRAL
ECG
IBP
NIBP
ALARMS
TRENDS
SELECT
EXIT
ADMIT
LEAD
SIZE
ZERO 1
ZERO 2
INTERVAL
VOLUME
LIMITS
DISPLAY
BEEP
VOLUME
HOLD 3 SEC. TO CLEAR
+
-
MUTE
DEFLATE
START
FREEZE
DISCHARGE
SILENCE
CO2 PUMP
HOLD 3 SEC. TO STOP
R
Datascope
Datascope
Gas Module SE™
41
42
43
44
45
46
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...