Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 131
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
1
In an unopened, undamaged package.
All Nellcor® accessories and sensors must be purchased form Nellcor® Inc., 25495
Whitehall Street, Hayward, Ca. 94545. To contact Nellcor®, call 1-800-NELLCOR.
1.6.10.2
Special Features
Automatic Calibration
The oximetry subsystem incorporates automatic calibration mechanisms. It is automatically
calibrated each time it is turned on, at periodic intervals thereafter, and whenever a new
sensor is connected. Also, the intensity of the sensor’s LEDs is adjusted automatically to
compensate for differences in tissue thickness.
Each sensor is calibrated when manufactured; the effective mean wavelength of the red LED
is determined and encoded into a calibration resistor in the sensor plug. The instrument’s
software reads this calibration resistor to determine the appropriate calibration coefficients
for the measurements obtained by that sensor.
Oximetry Operating Modes
The Passport’s three operating modes for SpO
2
enable it to make accurate measurements
despite differing levels of patient activity. In all three modes, the Passport updates its
measurements with every pulse beat. Data from the most recent beat replaces data from the
earliest beat and new averages are determined and displayed. The three modes are: 1
(default setting), 2, and 3.
1.
The default operating mode, uses a 5 to 7 second averaging time and is useful in
situations in which the patient is relatively inactive.
TABLE 1-1 Nellcor® SENSOR FAMILY
SELECTION
GUIDE
D25/D25L
ADULT
R-15 ADULT
N-25
NEONATAL
I-20
INFANT
D-20
PEDIATRIC
RS-10
ADULT
Patient Size
>30 kg
>50 kg
<3 kg >40 kg
1-20 kg
10-50 kg
>40 kg
Duration of
Use
Short or Long
Term
Short or Long
Term
Short or Long
Term
Short or
Long Term
Short or Long
Term
Short Term
Sterility
Sterile1
Sterile1
Sterile1
Sterile1
Sterile1
Non-sterile
Patient
Activity
Limited Activity
Inactive
Limited
Activity
Limited
Activity
Limited Activity
Limited
Activity
OXISENSOR
adult digit
oxygen
transducer
OXISENSOR
adult nasal
oxygen
transducer
OXISENSOR
neonatal
oxygen
transducer
OXISENSOR
infant digit
oxygen
transducer
OXISENSOR
pediatric digit
oxygen
transducer
RS-10
reflectance
oxygen
transducer
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...