Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 167
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
GM: ET AGENT
LOW
Appears when the End Tidal Anesthetic Agent
measurement is less than or equal to the value
set for the ET Agent Low Alarm.
Check patient.
GM: INSP AGENT
HIGH
Appears when the Inspired Agent
measurement is greater than or equal to the
value set for the INSP Agent High Alarm.
Check patient.
GM: INSP AGENT
LOW
Appears when the Inspired Agent
measurement is less than or equal to the value
set for the INSP Agent Low Alarm.
Check patient.
GM: CO
2
UNCALIBRATED
Appears after an unsuccessful calibration
attempt of the CO
2
sensor. The numeric data
for CO
2
will appear as - - -, and the CO
2
waveform will be a flatline.
Ensure proper gas
mixture is attached
tightly and regulator is
on. Repeat calibration
procedure. If problem
persists, call for service.
GM: O
2
UNCALIBRATED
Appears after an unsuccessful calibration
attempt of the O
2
sensor. The numeric data for
O
2
will appear as - - -, and the O
2
waveform
will be a flatline.
Ensure proper gas
mixture is attached
tightly and regulator is
on. Repeat calibration
procedure. If problem
persists, call for service.
GM: N
2
O
UNCALIBRATED
Appears after an unsuccessful calibration
attempt of the N
2
O sensor. The numeric data
for N
2
O will appear as - - -, and the N
2
O
waveform will be a flatline.
Ensure proper gas
mixture is attached
tightly and regulator is
on. Repeat calibration
procedure. If problem
persists, call for service.
GM: AGENTS
UNCALIBRATED
Appears after an unsuccessful calibration
attempt of the Agent sensor. The numeric data
for all agents will appear as - - -, and the agent
waveform will be a flatline.
Ensure proper gas
mixture is attached
tightly and regulator is
on. Repeat calibration
procedure. If problem
persists, call for service.
GM: FAILED
Appears when the Gas Module detects an
unrecoverable error in its own operation.
Call for Service.
GM:
DISCONNECTED
Appears when the Passport cannot detect
signals being sent by the Gas Module.
Ensure Gas Module is
turned on and interface
cable is properly
connected. If problem
persists, call for service.
SAMPLING ERROR
Appears when a sampling error occurs on one
or more Gas Module channels during
calibration.
Repeat calibration
procedure. If problem
persists, call for service.
CALIBRATION NOT
STARTED
Appears when the Gas Module is unable to
initialize calibration.
Repeat calibration
procedure. If problem
persists, call for service.
CALIBRATION
ERROR SAMPLING
ERROR
Appears when a sampling error occurs in all
four Gas Module channels during calibration.
Repeat calibration
procedure. If problem
persists, call for service.
CALIBRATION
ERROR ZEROING
ERROR
Appears when the Gas Module cannot
perform a Zero during calibration.
Repeat calibration
procedure. If problem
persists, call for service.
MESSAGE
REASON
SOLUTION
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...