Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 33
Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
For Lead II ECG Monitoring
1.3.6.2
Neonatal Electrode Placement
FIGURE 1-19
Neonatal Electrode Placement
Lead placement on the neonate with a three lead set is directed towards obtaining the best
respiration waveform. The thoracic impedance is measured between the RA and LA
electrodes therefore, the electrodes must be placed across the chest from each other to
optimize measurement of chest movement.
FIGURE 1-17
3-Lead Electrode Placement
(standard configuration)
For Lead II ECG Monitoring
• Place black electrode on left shoulder under
clavicle.
• Place white electrode on right shoulder
under clavicle.
• Place red electrode on lower left abdomen
under the sixth rib.
• Place lead select on lead II.
FIGURE 1-18
3-Lead Electrode Placement
(for MCL)
For MCL
• Place white electrode on left shoulder
under clavicle.
• Place black electrode on right sternal
border, 4th intercostal space.
• Place red electrode on midaxillary line, 5th
intercostal space.
• Select lead I for monitoring.
• For MCL, select lead II for monitoring.
RA
LA
LL
Red
Whit e
Blac k
L L
LA
R A
W hite
Red
Black
RA
LA
G
Red
Black
White
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...