Operation (XG)
Operation (5-Lead, 5L, LT and XG)
1 - 154
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
2.
After the “DIAGNOSTICS IN PROGRESS” message is displayed, press and hold the
FREEZE key (9) until a second beep is heard (approximately 2 seconds). The User
Configuration Mode will display.
The operation of the menu system is the same as the operation of the menu system during
normal operation (see “Use of Menus” on page 1-109). To access normal operation when
user configuration is complete, either time-out (no Set-Up key pressed within 1 minute) or
press the EXIT (3) key for 3 seconds. The following table describes the User Configuration
Mode menu structure:
The Trend Trigger setting is what causes new data to be stored in the trend memory. It may
be set to trigger whenever there is an alarm, an NIBP measurement is performed, the trend
timer expires (interval), or may be set to trigger at any combination of these items. The trend
interval is only used when trend is triggering on interval, and it is used to set the time
between interval triggers. The trend interval is independent of the NIBP interval.
USER CONFIG.
MENU
MENU ITEM
CHOICES
Date/Time
Year
0 to 99
Month
1 to 12
Day
1 to 31
Hours
0 to 23
Minutes
0 to 59
Trend
Trigger
Alarms, NIBP*, Interval, NIBP & Alarms,
Interval & Alarms, Interval & NIBP, Interval &
NIBP & Alarms
Interval
1, 2.5, 5*, 10, 20, 30, 60, 120 mins.
Temperature
Scale
Fahrenheit*, Centigrade
HR SpO
2
Size
Size Change
HR Large/SpO
2
Small, HR Small/SpO
2
Large*
Alarm Audio Delay
Delay
Off*, 4, 6, 8 sec.
Audio Alarm Standby
Function
Off*, On, On and Active on Powerup
Serial Output Type***
Serial Output Type
VISA*, VISA with Admit, Accutorr, Message,
DIAP, Gas Monitor
Accutorr Baud Rate
1200bps*, 2400bps, 4800bps, 9600bps
CO
2
Barometric Pressure**
500 to 800 mmHg, 760 mmHg*
CO
2
Units
TORR, kPa, %
Color Settings
ECG Color
White*, Green, Red
NIBP Color
White*, Green, Red
SpO
2
Color
White, Green*, Red
Gases Color
White, Green*, Red
IBP1 Color
White, Green, Red*
IBP
2
Color
White, Green, Red*
Resp Color
White, Green, Red*
Temp Color
White, Green, Red*
Gas Monitor***
CO
2
Units
TORR, mmHg, kPa, %
*
Initial Factory settings
**
On units with sidesteam CO
2
capability, there is automatic compensation for barometric pressure.
*** Only applicable if J1 connector is installed.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...