Operation (5-Lead, 5L, LT)
Operation (5-Lead, 5L, LT and XG)
1 - 54
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
For a graphic representation of the configuration menus see sections 1.3.19.11 through
1.3.19.19.
*
Factory default settings.
**
kPa is not available when VISA is selected as the serial output type.
The Trend Trigger setting is what causes new data to be stored in the trend memory. It may
be set to trigger whenever there is an alarm, an NIBP measurement is performed, the trend
timer expires (interval), or may be set to trigger at any combination of these items. The trend
interval is only used when trend is triggering on interval, and it is used to set the time
between interval triggers. The trend interval is independent of the NIBP interval.
The temperature scale can be changed between Fahrenheit and Centigrade.
The HR SpO
2
size change option allows the user to select a large SpO
2
reading compared
to HR or vice versa.
The Alarm Audio Standby allows the user to enable (or disable) this feature (see section
3.13.3).
The Serial Output Type allows the user to select the communication protocols for interfacing
with other specialized equipment: VISA, ACCUTORR (sends data the same way as an
ACCUTORR), message (for diagnostic purposes) with the Passport Monitor.
Once the unit is in the User Configuration Mode, the time and/or date can be changed using
the set-up keys as described in Section 1.3.4, “Use of Menus”.
1.3.15
Recorder (optional)
The Datascope Passport Recorder can provide a permanent record of a patient’s: systolic
pressure, diastolic pressure, mean pressure, heart rate, SpO
2
, CO
2
, respiration and
temperature. It is a two trace thermal strip chart recorder with an integral paper spool. The
recorder uses plain white thermal paper 5 cm wide (see Section 4.7 for replacement
instructions).
All grid patterns and data are printed by the recorder.
HR SpO
2
Size
Size Change
HR Large/SpO
2
Small, HR Small/
SpO
2
Large*
Alarm Audio Delay
Delay
Off*, 4, 6, 8 sec.
Audio Alarm
Standby
Function
Off*, On
Serial Output Type
Protocol
VISA*, ACCUTORR, Message,
DIAP
TABLE 1-10 User Configuration Menu List
USER CONFIG.
MENU
MENU ITEM
CHOICES
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...