Diagnostics and Calibration Procedure
Calibration
7 - 18
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
7.5.9
SpO
2
Diagnostics (Datascope SpO
2
Installed)
When this diagnostic is selected the Main Menu will be covered by the SpO2 Diagnostic
Menu which offers four different choices as shown in Figure 7-13. These tests verify the sub-
system of the SpO2 module as well as the accuracy of the selected SpO2 sensor.
FIGURE 7-9
Main Diagnostics Menu
7.5.9.1
Accuracy Test
The purpose of this test is to verify that the SpO2 Board performs accurately with any
selected SpO2 sensor. The test sends the infrared light output signal through both the red and
infrared channels. This produces a 1:1 ratio which corresponds to an SpO2 value of 85%.
The successful completion of this test signifies that the SpO2 Board can transmit and receive
light through the monitored sight, processing unit can successfully display the parameters. A
display of 85% means that the SpO2 Board can accurately display SpO2 over the entire
monitoring range.
a.
Highlight the Accuracy Test from the SpO2 Diagnostic Menu and press the SELECT
key to start the test.
b.
Shortly after the sensor is placed on a finger the SpO2 value is displayed along with
the operator’s pulse rate.
Accuracy Test Specification
% SpO2 = 85%
Press the RECORD key to obtain a printout of the test results.
c.
Press the END key to return to the SpO2 Test.
SERVICE DIAGNOSTICS
Error Journal
Exception Stack Frame
Communications
Recorder
Voltlage
Display
Keyboard
SpO2
Use arrow keys to move cursor, SELECT to choose a menu item
Press & hold the END key for 3 sec. to leave Service Diagnostics
NIBP: Idle
SpO2 Diagnostics
NIBP
Monitor Version:x.x.x:xxxx
PCM Version:x.x.x:xxx
SpO2 Version:x.x.x:xxxx
Thu May 14 13:07 1992
NIBP Version: x.x.:xxxx
NIBP DIAGNOSTICS
Accuracy Test
Sensor Test
DAC Test
Shared RAM Test
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...