Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
7 - 27
Calibration
Sidestream Pump Calibration
7.14.1
Sidestream CO
2
Pump Verification / Calibration Procedure
This test is to verify the proper flow rate of the Sidestream CO
2
Pump.
1.
Turn power switch to the on position and wait for the unit to power up in the normal
operating mode.
2.
Assemble the CO
2
Sidestream pump calibration parts as shown in Figure 7-13 and
attach to the Passport monitor.
3.
Verify zero cc/min, ±1 cc/min on flow meter before turning the CO
2
pump on. Press the
CO
2
pump key and measure the flow rate of the pump. Allow at least two minutes for
the flow meter to settle before any adjustment is made.
4.
Adjust VR2 on the CO
2
Control Module P/N 0670-00-0118 to achieve a flow rate of
180 cc/min, ±20 cc/min. Refer to Figure 7-14.
Occlusion Test:
1.
To simulate an occlusion, take a pair of Hemostats and crimp the tubing at the intake
port. An OCCLUSION message should appear in the lower left window.
FIGURE 7-13
Sidestream CO
2
Pump Calibration Kit
REQUIRED PARTS FOR SIDESTREAM CO2 PUMP CALIBRATION
ITEM
PART NUMBER
DESCRIPTION
1
0103-00-0443-01
Airway Adapter with Tubing
2
0008-00-0259
Nafion Tubing Assembly
3
0683-00-0405-11
Capillary Line 5 Feet
4
0103-00-0445-01
Filter
TO PASSPORT PUMP INPUT
FLOW
3
1
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...