Operation (5-Lead, 5L, LT)
Operation (5-Lead, 5L, LT and XG)
1 - 28
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
Suspension of Automatic NIBP Feature
To suspend an automatically timed measurement sequence or to end a measurement cycle
already in progress (deflate cuff):
1.
Press DEFLATE (12).
To take an immediate measurement and resume a suspended timed measurement
sequence:
2.
Press START (10).
NOTE:
Press DEFLATE (12) at any time to postpone a scheduled
measurement or to terminate a measurement cycle already
in progress.
CAUTION:
Observe extreme caution on all patients (neonates,
pediatrics, and adults) when NIBP is set to the Continuous
Mode. A 5 minute limit is placed on continuous
measurements. After 5 minutes the unit will take a
measurement once every 5 minutes. Reports have been
made of nerve injury occurring during use of automatically
cycled blood pressure cuffs. See the Appendix, “Precautions
when Using Automatically Cycled Blood Pressure
Cuffs”.Observe extreme caution on all patients (neonates,
pediatrics, and adults) when NIBP is set to the Continuous
Mode. A 5 minute limit is placed on continuous
measurements. After 5 minutes the unit will take a
measurement once every 5 minutes. Reports have been
made of nerve injury occurring during use of automatically
cycled blood pressure cuffs. See the Appendix, “Precautions
when Using Automatically Cycled Blood Pressure Cuffs”.
1.3.5.3
NIBP Pressure Limit Fail Safe
If the cuff pressure is over-pressurized, the cuff will automatically vent to atmosphere and the
NIBP message window reads CUFF OVERPRESSURE.
The unit must be turned off and back on again to reset the overpressure switch before any
new measurements are taken.
1.3.5.4
Cuff Inflation Time
If the cuff pressure does not attain 20 mmHg within 40 seconds of the start of inflation or if
the target pressure is not reached within another 60 seconds, then the cuff is vented and the
“RETRY” or “UNABLE TO MEASURE” message will display in the NIBP message window.
1.3.5.5
START and DEFLATE Functions
The START and DEFLATE functions have the following effects on the timed measurement
sequence.
• INTERVAL is set and you Press START (10):
An unscheduled measurement is made. Taking this unscheduled measurement does not affect
the timing of the interval cycle, therefore, the scheduled measurements will be taken as if
there were no interruptions. Only one measurement is taken for each measurement cycle -
even if the unscheduled measurement coincides with the scheduled measurement.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...