Power-Up Verification
Calibration
7 - 20
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
Memory Test 3 (Address Test): The main processor will write the entire memory address to
their corresponding memory location, then read them back to verify that the address bus is
not cross taking.
a.
Highlight one of the memory test from the SpO2 shared RAM test menu and press
the SELECT key to start the test.
b.
Each test is continuously performed until the END key is pressed. The completion
pass number is constantly updated on the display. The test is halted when a failure
occurs and address and write/read data patterns are displayed.
c.
Press the RECORD key to obtain a printout of the test results.
d.
Press the END key to return to halt the test.
7.5.10
SpO2 Diagnostics (Nellcor SpO
2
Installed)
The purpose of this test is for factory testing only. This test is not intended for field use.
7.5.11
SpO2 Diagnostics (Masimo SpO
2
Installed)
When this diagnostic is selected, the service menu will be replaced by a textual display of all
SpO2 errors recorded by the monitor. All errors will be automatically printed out on the
recorder as they are displayed on the screen, provided the unit is equipped with a recorder.
The DOWN ARROW key causes the display to clear, and the next page of messages to be
displayed. Continue to press the DOWN ARROW key unitl the message “Journal Empty” is
displayed.
NOTE:
There is no buffering of the error journal messages,
therefore once an error message is displayed, it will be lost
if the DOWN ARROW or END key is depressed.
The Masimo Diagnostic Journal is not intended for interpretation in the field. There are
hundreds of different messages which could be recorded, and the messages vary with
software revision. Also, not all recorded messages indicate an error. If you suspect a problem
with the unit, the Masimo Diagnostic Journal should be printed out to the recorder (if
installed) and forwarded to a Datascope Service Representative or Datascope Technical
Support department.
7.6
Power-Up Verification
1.
Plug the Power Supply Module into an appropriate power source and connect the D.C.
plug into the right side panel of the Passport.
2.
Turn on the Passport. Check for the following to occur:
• The LCD display is bright blue for approximately 5-8 seconds, followed by the
message:
Diagnostic in Progress
Selftest Complete
• At the end of the download, verify unit sets up display screens for signal inputs.
NOTE:
If no display is observed when unit is powered up, adjust
front panel “contrast ” control.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...