D. Establishing CO
2
/ CAPNOSTAT*
1.
Plug sensor into the side of the monitor.
2.
If a new CAPNOSTAT is used, calibrate the system.
3.
Apply sensor to the breathing circuit.
4.
Set Waveform 2 or 3 to CO
2
.
5.
Set the Resp. Source to CO
2
.
*
CAPNOSTAT is a trademark of NOVAMETRIX, Inc.
E. Recording Information
1.
Select wave to be recorded through Record Menu or if desired to record tabular trend,
press TREND/RETURN.
2.
Press Record to start recording function.
3.
Press Record again to stop the recording function.
F. Temperature Measurement
1.
Attach the desired temperature probe to the Temperature connector on the side of the
Passport. The Passport automatically detects which probe is connected, 400 or 700
series.
Detailed Operating Instructions
This section of the Service Manual provides guidelines and step-by-step instructions for
proper operation of the monitor. Numbers in parentheses ( ) relate to the displays and
controls described in “Controls, Indicators and Connectors (5-Lead, 5L, LT)” on page 1-3
1.3.1
Setting-up / Turning Power On
1.
Set the POWER switch (34) to OFF.
2.
Attach any peripheral equipment, i.e., Visa Central Station, Remote Color Display.
CAUTION:
Always attach all desired peripheral equipment prior to
powering on the Passport. If required to attach more or
change what is connected, turn the Passport off, attach
what is needed, and then turn the Passport on again.
3.
Attach the power pack to the DC POWER INPUT connector (32). If battery operation is
required, ensure that two fully charged batteries are installed.
NOTE:
If power pack is connected ensure that the screw is
tightened so that the cord does not detach.
4.
Set the POWER switch (34) to ON.
Internal self tests will run and the display will come on. A “DIAGNOSTIC IN
PROGRESS” message will display and once the self tests have been completed a “SELF
TEST COMPLETE” message will display. After this the main screen is displayed.
If any failures occur, see “Status Messages” on page 1-57 for further instructions.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...