Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 165
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
GM: MIXED
AGENTS
Appears in Inverse Video when more than one
anesthetic agent is detected by the system.
Message will disappear
when a single agent is
detected again.
GM: AIR LEAK
Appears when the system detects a pneumatic
leak. Also may appear when the Gas Module
has been turned on without a sample line
attached or if the Gas Module has been on for
a long period of time without the Passport
Monitor being on.
Turn Gas Module and
Passport off. Install/
check sample lines,
filters, water trap and
electrical connections.
Turn on Gas Module
and Passport Monitor.
GM: REPLACE TRAP
Indicates residue build-up on the water trap
membrane. This decreases air flow.
Replace water trap
reservoir.
GM: OCCLUSION
Appears when the system detects an
obstruction in the sampling line or the water
trap bottle is full.
Empty and rinse water
trap. Change water trap
if necessary. Check
sampling line and filter
for blockage, clear
sampling line if
possible. Replace
sampling line and/or
filter if necessary. If
problem persists call for
service.
GM: ZERO IN
PROGRESS
Appears when the system is zeroing all of it’s
channels. This appears whether initiated by
the user or is automatic.
This is Normal
Operation. Wait for
message to clear.
GM: CO
2
ZERO
ERROR
Appears when the system has been unable to
successfully zero the CO
2
sensor.
Manually start zeroing
the system again. If
problem persists call for
service.
GM: O
2
ZERO
ERROR
Appears when the system has been unable to
successfully zero the O
2
sensor.
Manually start zeroing
the system again. If
problem persists call for
service.
GM: N
2
O ZERO
ERROR
Appears when the system has been unable to
successfully zero the N
2
O sensor.
Manually start zeroing
the system again. If
problem persists call for
service.
GM: AGENT ZERO
ERROR
Appears when the system has been unable to
successfully zero the Anesthetic Agent sensor.
Manually start zeroing
the system again. If
problem persists call for
service.
GM: PUMP OFF
Appears when the system has turned off the
pump due to a pneumatic error.
Restart pump from Gas
Module menu. Refer to
“Gas Module Menu” on
page 1-182. If problem
persists, call for service.
GM: AGENT
MISMATCH - HAL
Appears in Inverse Video when the system
detects Halothane as the primary agent and
the manually selected agent is neither
Halothane nor the detected secondary agent.
Match the Agent
administered with the
Agent selected, or select
Agent Auto ID.
GM: AGENT
MISMATCH - ISO
Appears in Inverse Video when the system
detects Isoflurane as the primary agent and
the manually selected agent is neither
Isoflurane nor the detected secondary agent.
Match the Agent
administered with the
Agent selected, or select
Agent Auto ID.
MESSAGE
REASON
SOLUTION
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...