Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 99
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
J. Setting Alarms
1.
To access the Alarms Limits menu press the Alarms LIMITS key.
2.
Using the UP and DOWN and SELECT keys set parameter limits as desired.
Press SELECT key again to enter the chosen limits.
3.
Press the LIMITS key again to access further alarm parameters (when the Gas Module
is installed).
4.
Press the EXIT key to return to the main menu.
Detailed Operating Instructions
This section of the Operating Instructions provides guidelines and step-by-step instructions for
proper operation of the monitor. Numbers in parentheses ( ) relate to the displays and
controls described in “Controls, Indicators and Connectors (XG)” on page 1-77.
1.6.1
Setting-up / Turning Power On
1.
Set the POWER switch (38) to OFF.
2.
Attach any peripheral equipment (i.e., Visa Central Station, Remote Color Display).
CAUTION:
Always attach all desired peripheral equipment prior to
powering on the Passport. If required to attach more or
change what is connected, turn the Passport off, attach
what is needed, and then turn the Passport on again.
3.
Attach the power pack to the DC POWER INPUT connector (36). If battery operation is
required, ensure that two fully charged batteries are installed.
NOTE:
If power pack is connected ensure that the screw is
tightened so that the cord does not detach.
4.
Set the POWER switch (38) to ON.Internal self tests will run and the display will come
on. A “DIAGNOSTICS IN PROGRESS” message will display and once the self tests have
been completed a “SELF TEST COMPLETE” message will display. After these messages
the main screen is displayed.
If any failures occur, see “Status Messages” on page 1-158 for further instructions.
1.6.2
Factory - Default Control Settings
The following are the factory initial (default) control settings. These are the settings that the
unit will power up in unless a “current configuration” has been saved from the set up menu.
Changes can be made and saved to these default settings. The default settings can also be
restored. Refer to the “Set-up Menu” on page 1-176.
MENU
FUNCTION
DEFAULT
SETTINGS
Patient
Size
Current setting
Admit*
No
Discharge*
No
Full Disclosure*
Off
Arrhythmia*
Off
*
This item only appears if the serial output type is set to Visa or Visa w/ Admit.
**
These items only appear if the related function is installed.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...