Operation (5-Lead, 5L, LT)
Operation (5-Lead, 5L, LT and XG)
1 - 18
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
TABLE 1-3 Menu List
MENU
LIST
MENU ITEM
CHOICES
Patient
Size
Adult, Pediatric, Neonate
Clear Trend Memory
No, Yes
Pacer Enhancement
On, Off
Setup
Speed
12.5, 25, 50 mm/sec.
Resp Source
Off, ECG, CO
2
Resp Speed
3.125, 6.25, 12.5, 25 mm/sec.
Waveform 2
BP1, ECG casc., Pleth., Resp., CO
2
Waveform 3
BP1, BP2, Pleth., Resp., CO
2
Power up Settings
No Change, Save Current, Restore Factory
Record
Record on Alarm
Yes, No
Wave Selection
ECG, Pleth., Resp., ECG and BP1, ECG and BP2, BP1
and BP2, ECG and Resp., CO
2
Record Destination
Local, Remote Both
Alarm
HR
Low (Off, 30-100), High (Off, 100-250) bpm
SpO
2
Low (50-99), High (Off, 80-100) %
IBP1 Sys
Adult: Low (Off, 5 - 130), High (Off, 70 - 240) mmHg
Ped.: Low (Off, 5-130), High (Off, 40 - 180) mmHg
Neo.: Low (Off, 5-130), High (Off, 40 - 180) mmHg
IBP1 Dia
Adult: Low (Off, 5 - 90), High (Off, 40 - 130) mmHg
Ped.: Low (Off, 5 - 50), High (Off, 50 - 100) mmHg
Neo.: Low (Off, 5 - 50), High (Off, 50 - 100) mmHg
NIBP Sys
Adult: Low (Off, 50 - 150), High (Off, 70 - 240) mmHg
Ped.: Low (Off, 15 - 130), High (Off, 40-180) mmHg
Neo.: Low (Off, 15 - 130), High (Off, 40-180) mmHg
NIBP Dia
Adult: Low (Off, 30 - 120), High (Off, 40 - 130) mmHg
Ped.: Low (Off, 10 - 50), High (Off, 50 - 100) mmHg
Neo.: Low (Off, 10 - 50), High (Off, 50 - 100) mmHg
IBP2 (Mean)
Adult: Low (Off, 2 - 100), High (Off, 5 - 150) mmHg
Ped.: Low (Off, 2 - 50), High (Off, 5 - 100) mmHg
Neo.: Low (Off, 2 - 50), High (Off, 5 - 100) mmHg
Respiration
Low (Off, 5 - 50), High (Off, 30 - 200) bpm
Apnea Delay
Adult Off, 10 -40 secs.
Ped. Off, 10 - 30 secs.
Neonate 10 - 20 secs.
ETCO
2
Torr: Low (Off, 5 - 60), High (Off, 20 - 100) %: Low
(Off, 1 - 8), High (Off, 2 - 11) kPa: Low (Off, 1.0 -
8.0), High (Off, 2.0 - 11.0)
NIBP
Start Pressure
Adult: 100-260 mmHg
Ped.: 60-180 mmHg
Neonate: 40-120 mmHg
IBP
BP1 Scale Range
37.5, 75, 150, 300 mmHg
BP2 Scale Range
37.5, 75, 150, 300 mmHg
Resp
Scale
1, 2, 3, 4, 5
HR
HR Source
Auto, ECG, BP1, SpO
2
SpO
2
(Standard)
Pleth Size
1, 2, 3, 4, 5
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...