Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 53
Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
1.3.13
How to Set the Clock
1.
Enter the User Configuration Mode. (See Section 1.3.14.)
2.
Use the UP
p
and DOWN
q
arrow keys (1) to select Date or Time in the User
Configuration Menu. Press the SELECT Key (2) to choose that menu item. The current
choice is highlighted.
3.
Use the UP
p
and DOWN
q
arrow keys (1) to select a new setting. Once the desired
choice is highlighted, press the SELECT key (2).
4.
To save this setting press the SELECT key (2) and then the END key (3).
Repeat for each menu item that needs to change.
1.3.14
User Configuration Mode
The Trend Configuration, Time/Date, Temperature Scale, Heart Rate/SpO
2
Size Select,
Alarm Audio Delay, Audio Alarm Standby, Serial Output Type, and CO
2
settings may
changed in the User Configuration Mode. This function is only available at power up and not
during normal operation. The User Configuration Mode is accessed via a special power-up
sequence.
To enter the User Configuration Mode:
1.
Turn the POWER switch (34) ON.
2.
After the “DIAGNOSTIC IN PROGRESS” message is displayed, press and hold the
FREEZE key (6) until a second beep is heard (approximately 2 seconds). The User
Configuration Mode will display.
The operation of the menu system is the same as the operation of the menu system during
normal operation (See Section 1.3.4). To access normal operation when user configuration
is complete, either time-out (no Set-Up key pressed within 1 minute) or press the END (3) key
for 3 seconds. The following table describes the User Configuration Mode menu structure:
TABLE 1-10 User Configuration Menu List
USER CONFIG.
MENU
MENU ITEM
CHOICES
Date
Year
0 to 99
Month
1 to 12
Day
1 to 31
Time
Hours
0 to 23
Minutes
0 to 59
Trend
Trigger
Alarms, NIBP*, Interval, NIBP and
Interval, Interval and Alarms,
Interval and NIBP, Interval and
NIBP and Alarms
Interval
1, 2.5, 5*, 10, 20, 30, 60, 120
mins.
Temperature
Scale
Fahrenheit, Centigrade
CO
2
Barometric Pressure
500 to 800 mmHg
CO
2
Units
TORR, kPa**, %
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...