Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 163
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
CO
2
: ADAPTER CAL IN
PROGRESS
A CO
2
adapter calibration has been
requested and is in progress.
Wait for the adapter
calibration to complete.
It takes less than 30
seconds.
CO
2
: ZERO CAL FAILURE
The zero calibration was not completed
successfully. The zero calibration may
have been requested when the sensor
had not completed warming up.
Wait for the unit to
complete warming up.
Repeat the zero
calibration.
CO
2
: ADAPTER CAL
FAILURE
The adapter calibration was not
completed successfully. The adapter
calibration may have been requested
when the sensor had not completed
warming up or the sensor was not
correctly installed on the airway
adapter.
Wait for the unit to
complete warming up or
check the installation of
the sensor on the
airway adapter. Repeat
the adapter calibration.
CO
2
: CAL FAILED - INSERT
ZERO CELL
A zero calibration was attempted while
the sensor was not attached to the zero
cell.
Attach the sensor to the
zero cell and repeat the
zero calibration. The
zero cell is located on
the sensor cable.
CO
2
: CAL FAILED - INSERT
ADAPTER
An adapter calibration was attempted
while the sensor was attached to the
zero or reference cell.
Remove the sensor from
the zero or reference
cell and attach it to the
air-way adapter. Repeat
the adapter cal.
CO
2
: ZERO CAL
COMPLETE
The zero calibration has been
successfully completed.
Remove the CO
2
sensor
from the zero cell and
place back on the
adapter. Perform an
adapter zero.
CO
2
: ADAPTER CAL
COMPLETE
The adapter calibration has been
successfully completed.
Normal operation. The
message is removed
after a few seconds.
CO
2
: FAILURE
The CO
2
subsystem has failed.
Cycle the power on the
monitor to determine if
the problem persists. If
the message returns
then the unit needs to be
serviced.
MESSAGE
REASON
SOLUTION
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...