Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
7 - 7
Calibration
Diagnostics and Calibration Procedure
1.
Motor Pump Test
The purpose of this test is to determine if the output of the pump is adequate.
a.
Connect the dummy cuff/test chamber to the side panel luer fitting.
NOTE:
The dummy cuff/test chamber must be used for this test,
otherwise the test results are invalid.
b.
Highlight the Motor Pump Test option and press the SELECT key to start the test. The
pump will start and inflate the chamber to 250 mmHg, then pressure is released.
During the inflation/deflation cycle the current pressure is displayed in the pressure
window. The time required to reach the target pressure will be displayed in units of
xx.x seconds as shown in Figure 7-4.
Specification
Pump to 250 mmHg < 16.0 seconds
FIGURE 7-4
Motor Pump Test Display
2.
Bleed Rate Test
The purpose of this test is to determine if the solenoids and orifices are performing properly.
a.
Connect the dummy cuff/test chamber to the side panel luer fitting.
NOTE:
The dummy cuff/test chamber must be used for this test,
otherwise the test results are invalid.
NOTE:
The NIBP hose with a safety leak fitting cannot be used,
otherwise the test results are invalid.
b.
Highlight the Bleed Rate Test option and press the SELECT key to start the test.
Selecting this option causes the pump motor to inflate the test chamber to 170
mmHg. The largest orifice is then opened. The time required for the pressure to drop
from 150 mmHg to 130 mmHg is counted. The test is repeated for the middle and
smallest orifice. The chamber is then inflated to 250 mmHg. This time the dump valve
is opened and the time required for the pressure to drop from 250 mmHg to 20
mmHg is counted. The pressure is then released and the bleed rate for each orifice is
displayed in units of xx.x mmHg/sec. The current pressure is displayed in the
pressure window during the entire test. The format for displaying the test results is
shown in Figure 7-5.
SERVICE DIAGNOSTICS
Error Journal
Exception Stack Frame
Communications
Recorder
Voltlage
Display
Keyboard
SpO2
Use arrow keys to move cursor, SELECT to choose a menu item
Press & hold the END key for 3 sec. to leave Service Diagnostics
NIBP: Idle
SpO2 Diagnostics
NIBP
Monitor Version:x.x.x:xxxx
PCM Version:x.x.x:xxx
SpO2 Version:x.x.x:xxxx
Thu May 14 13:07 1992
NIBP Version: x.x.:xxxx
NIBP DIAGNOSTICS
Pneumatic
Pulse Channel
Pressure Calibration
NIBP PNEUMATIC TESTS
Motor Pump Test
Bleed Rate Test
Leak Test
Overpressure Test
Time required to pump to 250 mmHg - 10.7 sec.
Press Deflate to end test.
CUFF:
O
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...