Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
7 - 33
Calibration
Special Options Menu
Notch Filter:
1.
Press the Select key to get to the Notch Filter menu. Press the UP or DOWN key to
enable or disable the Notch Filter. when the Notch Filter is disabled the 50 hz or 60 hz
frequency filter is not active.
ESU Filter:
1.
Press the Select key to get to the ESU Filter Menu. Press the UP or DOWN key to disable
the ESU Filter or place the filter in the Auto mode. When the ESU Filter is disabled the
Electrosurgical Unit filter is not active.
Pump Status (XG Sidestream units only):
1.
This menu is non-selectable. Pump Status displays the status of the Sidestream CO
2
in
Bytecode.
NOTE:
Pump status is not intended for interpretation in the field. If
a problem is suspected with CO
2
pump operation, copy the
pump status bytecode and forward to Datascope Technical
Support Department.
Pump Current (XG Sidestream units only):
1.
This menu is non-selectable. Pump Current displays the amount of current drawn by the
Sidestream CO
2
pump assembly when activated.
Barometric Pressure (XG Sidestream units only):
1.
This menu is non-selectable. Barometric Pressure is displayed in mmHg’s and is updated
as per real time pressure.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...