Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 171
Operation (5-Lead, 5L, LT and XG)
Operation (XG)
No Resp.
Waveform or
Rate Displayed
-Patient not connected to a patient safety cable.-
Respiration parameter is “OFF”.
-CO
2
/Off and no CO
2
.
-Patient connected using Patient ESIS Choke/Cable.
-Turn respiration on
(“OFF” will be
displayed in resp.
window).Check that
proper patient cable is
used.-Use 3-lead Patient
Cable - non ESIS.
Waveform
Acquisition
Failure
-Front End Failure-External Supply Failure
-+12/-23 Volt Board Failure-CPU Board Failure
-CO
2
Interface Board Failure (optional)
-Replace Front End
Board.
-Check External Supply
Output voltage.
-Replace F1 fuse on
+12/-23 Volt Board.
-Replace CPU Board.
-Replace CO
2
Interface
Board
PROBLEM
REASON
SOLUTION