• Incorrect placement can also reduce the acquired sensor signal, and therefore
compromise performance. Select an alternate site (toe) or use a FLEXISENSOR® SD if the
sensor can not be placed on the patient’s finger correctly or if the fingernails interfere
with the acquisition of a reliable signal.
• Use of the DATASENSOR is not recommended for long-term monitoring (4-6 hours). For
monitoring situations exceeding 4-6 hours, either reposition the DATASENSOR every 2-4
hours to a different site (finger/toe) or use a FLEXISENSOR® SD with its appropriate
bandage.
• Do not over-tighten the sensor bandages. Excessive pressure on the monitoring site can
affect SpO
2
readings and may reduce readings below true SpO
2
. Excessive pressure can
also result in pressure necrosis and other skin damage.
• Sensor configuration provides uninterrupted monitoring in the following situations:
Electro-cautery Noise - ESU rejection is designed into the sensors.
Motion Artifact - The monitor’s software adjusts the “averaging period” increasing it
during motion and reducing it during inactivity. This decreases the number of monitoring
interruptions and false alarms.
Weak Peripheral Pulses - The monitor’s gain is automatically increased to track pulses on
patients with decreased peripheral perfusion.
• SpO
2
measurements may interfere with Magnetic Resonance Imaging (MRI) procedures.
• SpO
2
is calibrated to display functional saturation.
*
Non-adhesive bandages are recommended for premature infants to minimize prenatal skin damage.
**
See Accessories, Chapter 5, for more detailed information
*** Additional choices: 0060-00-0026-02 (10’ sensor cable) 0020-00-0071-01 (3’ sensor cable plus 7’ extension
cable)
TABLE 1-7 Sensor Selections
SENSORS
LARGE
ADULT
(LA)
ADULT (A)
PEDIATRIC
(P)
INFANT
(I)
NEONATE
(N)
ADULT
EAR (AE)
DATA-
SENSOR
Approximate
Patient
Weight
>80kg/>176
lbs
30 -90kg/
66 - 198 lbs
10 - 40kg/22
- 88 lbs
4.5 - 10kg/
10 - 22 lbs
Up to 5kg/Up
to 11 lbs
>40kg/>88
lbs
40+ kg/90+
lbs
Where Used
Fingers, Toes
Fingers, Toes
Fingers, Toes
Feet, Palms,
Big Toes
Feet, Palms,
Heel,Calf
Adult Ear
Fingers, Toes
Long or Short
Term
Monitoring
Long and
Short Term
Long and
Short Term
Long and
Short Term
Long and
Short Term
Long and
Short Term
Long and
Short Term
Short Term
ESIS
Included
Included
Included
Included
Included
Included
Included
Reusable
Yes Up to 20
Uses
Yes Up to 20
Uses
Yes Up to 20
Uses
Yes Up to
20 Uses
Yes Up to 20
Uses
Yes Up to 20
Uses
Yes 6-Months
Bandage
Type
Adhesive,
Disposable
Adhesive,
Disposable
Adhesive,
Disposable
Non-
Adhesive*,
Disposable
Non-
Adhesive*
Disposable
N/A
N/A
**Part
Numbers
Sensors
Bandages
0998-00-
0076-06
0998-00-
0076-05
0998-00-
0076-04
0998-00-
0074-03
0998-00-
0074-04
0998-00-
0074-05
0600-00-
0026-01 (3"
sensor
cable)***
0683-00-
0409-01
0683-00-
0409-02
0683-00-
0409-03
0683-00-
0415
0683-00-
0440
N/A
N/A
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...