Passport 5-Lead, 5L, LT, XG Service Manual
0070-00-0420
1 - 71
Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
1.3.19.9
SpO
2
Menu / Nellcor SpO
2
Menu
Pleth Size - Sets the scale of the plethysmograph waveform.
Mode - Sets the scale of the plethysmograph waveform on units with Nellcor.
FIGURE 1-38
SpO
2
1.3.19.10
CO
2
Menu
CO
2
Scale - Select the size of the displayed CO
2
waveform.
Start Zero Calibration - Select Yes to initiate the calibration of the sensor.
Start adapter cal - Select Yes to initiate the calibration of the airway adapter with the sensor.
N
2
O Compensation - Select ON to compensate for interference caused by nitrous oxide
present in the gas being analyzed.
O
2
Compensation - Select ON to compensate for interference caused by oxygen present in
the gas being analyzed.
CHANGE SPO2
Pleth Size:
Choices: 1, 2, 3, 4
= Adjust valve
END = quit
SELECT = Enter/move
3
CHANGE SPO2
Mode:
Choices: 1, 2, 3
= Adjust valve
END = quit
SELECT = Enter/move