Operation (5-Lead, 5L, LT)
Operation (5-Lead, 5L, LT and XG)
1 - 40
0070-00-0420
Passport 5-Lead, 5L, LT, XG Service Manual
3.
The digital SpO
2
value will be displayed in the Parameter Window.
4.
If a pleth waveform is desired, enter the set-up menu as described in “Use of Menus” on
page 23 to display it as Waveform 2.
5.
Press the BEEP VOLUME (18) to set the volume of the SpO
2
beep.
1.3.9.1
Nellcor® Sensors
Nellcor® provides a family of sensors suitable for a wide variety of clinical settings and
patients. Specific sensors have been developed for neonates, infants, children, and adults.
OXISENSOR™ oxygen transducers are sterile adhesive sensors with optical components
mounted on adhesive tape. OXIBAND® oxygen transducers and the DURAFORM™ oxygen
transducer system are reusable sensors that are applied with disposable adhesive. The
DURASENSOR® DS-100A adult digit oxygen transducer is a reusable sensor with its optical
components mounted in a plastic casing. The Nellcor® RS-10 reflectance oxygen transducer
is an adhesive sensor for application to forehead or temple.
NOTE:
Nellcor®, OXIBAND® and DURASENSOR® are registered
trademarks of Nellcor® Incorporated. OXISENSOR™ and
DURAFORM™ are trademarks of Nellcor Incorporated.
Selecting a Sensor
Sensors are designed for specific sites on patients with designated weight ranges. To select
the appropriate sensor, consider the patient’s weight, level of activity, adequacy of
perfusion, which sensor sites are available, whether sterility is required, and the anticipated
duration of monitoring.
Cleaning and Re-use
Do not immerse any OXISENSOR™, DURASENSOR®, OXIBAND®, or DURAFORM™
oxygen transducer, the Nellcor® RS-10 oxygen transducer, or any Nellcor® adhesive in
water or cleaning solution. Clean DURASENSOR®, OXIBAND®, and DURAFORM™ oxygen
transducers, and the Nellcor® RS-10 oxygen transducer by wiping with a disinfectant such
as 70% alcohol. Do not sterilize by irradiation, steam, or ethylene oxide. Use a new
OXIBAND® adhesive wrap or FORM-A adhesive bandage for each patient. Do not re-
sterilize OXISENSOR™ oxygen transducers.
Performance Considerations
To ensure optimal performance, use an appropriate sensor, apply it as directed, and observe
all warnings and cautions.
If excessive ambient light is present, cover the sensor site with opaque material. Failure to do
so may result in inaccurate measurements. Light sources that can affect performance include
surgical lights, especially those with a xenon light source, bilirubin lamps, fluorescent lights,
infrared heating lamps, and direct sunlight.
Summary of Contents for Passport 5-Lead
Page 1: ...Service Manual 5 Lead 5L LT XG Datascope Passport 0070 01 0420 indd 1 4 11 11 4 45 PM...
Page 2: ...5 Lead 5L LT XG Service Manual Datascope Passport 0070 02 0420 indd 1 4 11 11 4 45 PM...
Page 324: ...5 4 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 502: ...6 100 0070 00 0420 Passport 5 Lead 5L LT XG Service Manual This page intentionally left blank...
Page 544: ...0070 00 0420 Rev T April 13 2011...