Operation (5-Lead, 5L, LT and XG)
Operation (5-Lead, 5L, LT)
To perform a Capnostat® sensor calibration:
1.
Verify the Passport is turned on and the Capnostat® is plugged in and warmed-up.
2.
Place the Capnostat® onto the ZERO cell.
3.
The “Sensor on Zero Cell” message is displayed.
4.
Select Start Zero Cal - Yes from the CO
2
menu. The “Zero Cal in Progress” message is
displayed. The "Zero Cal Complete" message is displayed when complete and a 0
value is displayed.
5.
Remove sensor from the zero cell and place onto the airway adapter.
FIGURE 1-23
Zero Cell
1.3.12
Alarms
The Datascope Passport Monitor provides high and low alarm limits for heart rate (HR),
systolic pressure (BP1/NIBP Sys), diastolic pressure (BP1/NIBP Dia), mean pressure (BP2
Mean), respiration rate, ETCO
2
, and SpO
2
. An alarm for apnea delay is also provided.